WonATech
   
 

 
Product | Impedance Analyzer
ZIVE SP5
Applicaton
  • Versatile high quality potentiostat/galvanostat/impedance analyzer
  • Control voltage range : ±10V, ±1V, ±100mV
  • Control current range : 50pA ~ 5A (500pA with gain, 11 ranges)
  •  
       
      Hardware

    Power amplifier (CE)

    Power

    40Watt

    Compliance voltage

    ±10V

    Max. current

    ±5A within power

    Control speed selection

    8 ea

    Bandwidth

    1MHz

    Slew rate

    10V/usec

    Potentiostat mode(voltage control)

    Voltage control

    Control voltage range

    ±10, ±1V, ±100mV

    Voltage resolution

    16 bit per each range

    Voltage accuracy

    ±0.02% fs(gain x1)

    Max. scan range

    ±10V vs. ref. E

    Current measurements

    Current range

    11 ranges (auto/manual setting)
    5nA to 5A (500pA with gain)

    Current resolution

    16 bit;
    150uA, 15uA, 1.5uA, 1500nA, 15nA,
    1.5nA, 150pA, 15pA, 1.5pA, 150fA, 15fA

    Current accuracy

    ±0.02% f.s.(gain x1)>500nA

    Galvanostat mode(current control)

    Current control

    Control current range

    max. ±5A
    ± full sacle depending on selected range

    Current resolution

    16 bit;
    150uA, 15uA, 1.5uA, 1500nA, 15nA,
    1.5nA, 150pA, 15pA, 1.5pA, 150fA, 15fA

    Current accuracy

    ±0.02% f.s.(gain x1)>500nA f.s.

    Voltage measurements

    Voltage range

    10V, 1V, 100mV

    Voltage resolution

    16 bit; 0.3mV, 30uV, 3uV

    Voltage accuracy

    ±0.02% fs(gain x1)

    Electrometer

    Max. input voltage

    ±10V

    Input impedance

    2x1013Ω||4.5pF

    Bandwidth

    >22MHz

    CMRR

    >114dB

    EIS(Internal FRA) for system

    Frequency range

    10uHz ~ 1MHz

    Frequency accuracy

    <0.01%

    Frequency resolution

    5000/decade

    Amplitude

    0.1mV ~ 5Vrms(potentiostatic)
    0.1 ~ 70% f.s.(galvanostatic)

    Mode

    Static EIS:
    potentiostatic, galvanostatic,
    pseudo-galvanostatic, OCP
    Dynamic EIS:
    potentiodynamic, galvanodynamic
    Fixed frequency impedance:
    potentiostatic, galvanostatic,
    potentiodynamic, galvanodynamic
    Multisine EIS:
    potentiostatic, galvanostatic
    Intermittent PEIS/GEIS

    Main system

    PC communication

    USB2.0 high speed

    Line voltage

    100~240VAC, 50/60Hz

    Size/Weight

    179x270x378.4mm(WxDxH) / 7.65kg

    Power consumption

    24V@6.3A, 150Watt

    LED

    run, busy, potentiostat, galvanostat

    Software

    Max. steps per experiments

    1000

    Shutdown safety limits

    voltage, current, temperatrue etc.

    Max. sampling rate

    2usec or 3usec depending on data point numbers

    Min. sampling time

    unlimited

    Sampling condition

    time, dv/dt, dI/dt, temp. etc

    Interface for system

    Auxiliary port

    - Digital output

    3 (open collector)

    - Digital input

    2 (photo coupler)

    - Auxiliary voltage inputs

    3 analog input: ±10V
    for measurement of WE vs. CE
    CE vs. RE or other signal

    - Analog output

    1 analog output : ±10V
    for stirrer, MFC, RDE etc.

    Misc.port

    - Signal generator output

    1 analog output for FRA output or
    waveform generation output

    - Peripheral communication

    12C to control external devices

    - Temperature measurement

    1 K-type thermocouple input

    - Zero resistance ammeter

    5nA ~ 5A ranges

    System

    Cell cable

    1 meter shielded type (standard)
    working, reference, counter, working sense

    Control
    DAC

    DSP with FPGA
    2x16bit DAC(50MHz) for bias & scan
    1x16bit DAC(1MHz) for analog output

    Data acquisition
    ADC

    2x16bit ADCs(500kHz) for voltage & current
    4x16bit ADCs(250kHz) for auxiliary reading
    and temperature reading

    Calibration

    Automatic

    Filter selection

    4 ea(5Hz, 1kHz, 500kHz, 5MHz)

    Scan rate

    0 ~ 200V/sec in common mode
    0 ~ 5000V/sec in fast mode

    Max. channel No.

    16 channels via USB connection

    Max. output power 40Watt

    Internal data memory

    542,000 points

    LCD display

    DC & EIS mode automatically

    PC requirement

    Operating system

    Windows 7/8/10(32bit/64bit OS)

    PC specification(minimum)

    Pentium4, RAM 1GB or higher

    Display

    1600x900 high color or higher

    USB

    high speed 2.0

    General

    Dummy cell

    One external dummy cell included

    Thermocouple

    1.5meter K-type (option)

    Impedance data analysis software

    ZMANTM software

    DC data analysis software

    IVMANTM software

    The specifications are subject to change without notice.

       
     
    WPG Series
    Mid power type WPG100S (Max 400watt)
     
    WBCS mid power type
    WBCS3000S
     
    ZIVE BP/MP Series
    ZIVE MP5
     
    ZMAN
    ZMAN, Impedance Data Analysis Software