단채널 임피던스 모니터 Zcon |
별도의 전자부하기 및 potentiostat/galvanostat를 사용하여 단채널 임피던스 실험을 할 수 있는 장비입니다. 일반 임피던스 실험 뿐만 아니라 고전류 실험이 가능하여 배터리, 연료전지, 슈퍼캐퍼시터, 부식 등의 응용에 적합합니다.
|
|
|
|
|
|
|
|
|
응용 분야
|
|
|
Analog Out |
as Signal Generator |
number of channels
|
1
|
configuration
|
single-ended |
max. output |
-11.0 to +11.0 V (DC + AC) |
voltage offset |
<0.5mV, software corrected zero |
dc bias |
range |
resolution |
|
0.0 to 5.0 V |
0.076 mV |
|
0.0 to +10.0 V |
0.153 mV |
|
-5.0 to +5.0 V |
0.153 mV |
|
-10.0 to +10.0 V |
0.305 mV |
|
-2.5 to +2.5 V |
0.076 mV |
|
-2.5 to +7.5 V |
0.153 mV |
ac waveform |
|
pre-defined type |
dc, sine, cosine, ramp sawtooth, triangle, square, pulse, multi-tone |
frequency range |
1uHz to 100kHz resolution : 5000 steps/decade |
frequency accuracy |
typ. 75ppm, max.±200ppm
|
frequency stability |
<2 ppm @ 1 kHz |
|
<20 ppm @ 10 kHz |
|
<200 ppm @ 100 kHz |
|
<2000 ppm(0.2%)@ 1 MHz |
amplitude |
1 mVpp to 5 Vpp |
post-gain/attenuation |
-44 dB to +40 dB with 6 dB step, automatic gain selection |
reconstruction filter |
10 to 150 kHz 8th order low passfilter with 10 kHz step or by-pass |
gain error |
<0.5% |
Analog In |
as Frequency Analyzer |
number of channels |
1 for current input & 1 for voltage input |
configuration |
differential |
max.commonmodevoltage |
- Model Zcon : ±10 V
- Model ZconH : ±100 V
|
voltage offset |
<0.5 mV, software corrected zero |
bandwidth |
550 kHz |
input impedance |
110 kOhm |
pre-attenuation |
-20dB (x0.1) |
post-gain/attenuation |
-44 dB to +40 dB (x100) with 6 dB step or x200, x400, x800, x1600 |
anti-aliasing filter |
10 to 150 kHz 8th order low pass filter with 10 kHz step or by-pass
|
CMRR |
>80 dB @ 1kHz, >60 dB @ 10 kHz,, >40 dB @ 100 kHz (refer to the below graph) |
Expansion Ports |
|
12C in & out
|
reserved for future |
General |
|
interface
|
USB 2.0 high speed |
power
|
external 50W AC-DC adapters, +5/+15/-15 VDC with AC input of 100 to 240V, 2A, 50/60Hz |
operationcondition
|
0 to 50°C, 0 to 90% humidity (non-condensing) |
warranty
|
1 year parts and labor on defects in materials and workmanship |
The specifications are subject to change without notice. |
|
| |
|
|
|
|
|
|
|
|
|
|
|
|