Power amplifier (CE) |
|
Power
|
40Watt(40V@1A)
|
Compliance voltage
|
±40V |
Max. current |
±1A |
Control speed selection |
8 ea |
Bandwidth |
700kHz |
Slew rate |
7V/usec |
Potentiostat mode(voltage control) |
|
Voltage control |
|
Control voltage range |
±40, ±4V, ±400mV |
Voltage resolution |
16 bit per each range |
Voltage accuracy |
±4mV ±0.1% of setting |
Max. scan range |
±40V vs. ref. E |
Current measurements |
|
Current range |
11 ranges (auto/manual setting)
1nA to 1A (100pA with gain) |
Current resolution |
16 bit; 30uA, 3uA, 300nA, 30nA, 3nA, 300pA,
30pA, 3pA, 300fA, 30fA, 3fA |
Current accuracy |
±0.1% f.s.(gain x1)>100nA
|
Galvanostat mode(current control) |
|
Current control |
|
Control current range
|
max. ±1A ± full sacle depending on selected range
|
Current resolution
|
16 bit; 30uA, 3uA, 300nA, 30nA, 3nA, 300pA,
30pA, 3pA, 300fA, 30fA, 3fA
|
Current accuracy
|
±0.1% f.s.(gain x1)>100nA
|
Voltage measurements |
|
Voltage range
|
±40V, ±4V, ±400mV |
Voltage resolution
|
16 bit; 1.2mV, 120uV, 12uV
|
Voltage accuracy
|
±4mV ±0.1% of reading
|
Electrometer |
|
Max. input voltage
|
±40V |
Input impedance
|
2x1013Ω||1pF
|
Bandwidth
|
>2MHz
|
CMRR
|
>95dB
|
EIS(Internal FRA) for system |
|
Frequency range
|
10uHz ~ 600kHz
|
Frequency accuracy
|
<0.01%>
|
Frequency resolution
|
5000/decade
|
Amplitude
|
0.1mV ~ 20Vrms(potentiostatic) 0.1 ~ 70% f.s.(galvanostatic)
|
Mode
|
Static EIS: potentiostatic, galvanostatic, pseudo-galvanostatic, OCP Dynamic EIS: potentiodynamic, galvanodynamic Fixed frequency impedance: potentiostatic, galvanostatic, potentiodynamic, galvanodynamic Multisine EIS: potentiostatic, galvanostatic Intermittent PEIS/GEIS
|
Main system |
|
PC communication
|
USB2.0 high speed
|
Line voltage
|
100~240VAC, 50/60Hz
|
Size/Weight
|
179x270x378.4mm(WxDxH) / 7.65kg
|
Power consumption
|
48Watt
|
LED
|
run, busy, potentiostat, galvanostat
|
Software |
|
Max. steps per experiments
|
1000
|
Shutdown safety limits |
voltage, current, temperatrue etc. |
Max. sampling rate |
2usec or 3usec depending on data point number |
Min. sampling time |
unlimited |
Sampling condition |
time, dv/dt, dI/dt, temp. etc
|
Interface for system |
|
Auxiliary port |
|
- Digital output
|
3 (open collector)
|
- Digital input
|
2 (photo coupler)
|
- Auxiliary voltage inputs
|
3 analog input: ±10V for measurement of WE vs. CE CE vs. RE or other signal
|
- Analog output
|
1 analog output : ±10V for stirrer, MFC, RDE etc.
|
Misc.port
|
|
- Signal generator output
|
1 analog output for FRA output or waveform generation output
|
- Peripheral communication
|
12C to control external devices
|
- Temperature measurement
|
1 K-type thermocouple input
|
- Zero resistance ammeter
|
1nA ~ 1A ranges
|
System
|
|
Cell cable
|
1 meter shielded type (standard) working, reference, counter, working sense
|
Control DAC
|
DSP with FPGA 2x16bit DAC(50MHz) for bias & scan 1x16bit DAC(1MHz) for analog output
|
Data acquisition ADC
|
2x16bit ADCs(500kHz) for voltage & current 4x16bit ADCs(250kHz) for auxiliary reading and temperature reading
|
Calibration
|
Automatic
|
Filter selection
|
4 ea(5Hz, 1kHz, 500kHz, 5MHz)
|
Scan rate
|
0 ~ 200V/sec in common mode 0 ~ 5000V/sec in fast mode
|
Max. channel No.
|
16 channels via USB connection
|
Internal data memory
|
542,000 points |
LCD display
|
DC & EIS mode automatically
|
PC requirement
|
|
Operating system
|
WindowsXP SP3/7/8//10(32bit/64bit OS)
|
PC specification(minimum)
|
Pentium4, RAM 1GB or higher
|
Display
|
1600x900 high color or higher
|
USB
|
high speed 2.0
|
General
|
|
Dummy cell
|
One external dummy cell included
|
Thermocouple
|
1.5meter K-type (option)
|
Auxiliary cable
|
Option
|
Misc. cable
|
Option
|
Impedance data analysis software
|
ZMANTM software |
DC data analysis software
|
IVMANTM software |